Fluke 726 Precision Multifunction Calibrators

Key features

  • Precision multifunction, temp, frequency, voltage, and mA calibrator
  • Calibrates thermocouples and RTD temperature instruments
  • Stores up to eight calibration results in memory for analysis
  • Built-in 24 V Loop Power Supply
Category:

Fluke 726 Precision Multifunction Calibrators

Process calibration designed specifically for process industries
The Fluke 726 is a precise and powerful, yet easy-to-use, field calibrator. It features broad workload coverage, calibration power, and unsurpassed accuracy needed by process professionals. It measures and sources almost all process parameters to calibrate almost anything. Use it to test sensors and valves, and to test and calibrate transmitters. The 726 interprets results without the help of a calculator and stores measurement data for later analysis.

Measurement capabilities

This precision multifunction calibrator measures volts, mA, pressure, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters. The 726 also sources and simulates volts, mA, thermocouples, RTDs, frequency to calibrate transmitters. Measures and/or sources pressure using when paired with any of the 50 Fluke 750P series Pressure Modules to test and calibrate pressure transmitters and other devices. The 726 can also:

  • Source mA with simultaneous pressure measurement to conduct valve and I/P test
  • Conduct enhanced flowmeter testing with frequency totalizer and frequency pulse train source mode

Designed for ease of use

The 726 has built-in features that make it easier to use in nearly any measurement situation. From its ability to store frequently used test setups, to programming custom RTD probe calibration constants for enhanced temperature measurement, or the integrated voltage input protection design that improves reliability, the 726 is built to perform. The use of two separate channels even allows you to measure, source, or view process signals simultaneously. The backlit display provides better visibility in low-light conditions, and the standard 3-year warranty provides extra confidence that the 726 will be there when you need it to be.

You can also use the 726 to:

  • Capture the set, reset, and deadband of a pressure switch with the integrated switch test
  • Calculate the transmitter error % to interpret calibration results without the use of a calculator
  • Perform fast linearity tests with the auto-step and auto-ramp features
  • Power a transmitter during test using 24 V loop supply and simultaneous mA measurement

Specifications

Measurement Accuracy
Voltage DC
30.000 V
0.01% + 2 mV (upper display)
20.000 V
0.01% + 2 mV (lower display)
90.00 mV 0.01% + 10 μV
-10.00 mV to 75.00 mV
0.025% + 1 count (via TC connector)
Current DC 24.000 mA 0.01%+ 2 mA
Resistance 0.0 to 4000.0 Ω ± 0.015%
Frequency
2.0 to 1000.0 CPM
0.05% + 0.1 cpm
1.0 to 1000.0 Hz 0.05% + 0.1 Hz
1.00 to 15.00 kHz
0.05% + 0.01 kHz
Sensitivity
1 V peak-to-peak-minimum
Pressure Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC
100.00 mV 0.01% + 10 μV
20.000 V 0.01% + 2 mV
-10.00 mV to 75.00 mV
0.025% + 1 count (via TC connector)
Current DC
24.000 mA (Source) 0.01% + 2 μA
24.000 mA (Simulate) 0.01% + 2 μA
Resistance
Range, accuracy (percent of reading + floor) and excitation current
5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5 mA
5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3 mA
400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA
1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA
Frequency
2.0 to 1000.0 CPM 0.0005
1.0 to 1000.0 Hz 0.0005
1.0 to 10.0 kHz 0.0025
10.0 to 15.0 kHz 0.005
Waveform
5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure and source accuracy
Cu10 1.8°C
NI-120 0.15°C
PT-100 (385) 0.45°C
PT-100 (392) 0.35°C
PT-100 (391-JIS) 0.35°C
PT-200 (385) 0.95°C
PT-500 (385) 0.55°C
PT-1000 (385) 0.35°C
RTD Resolution 0.01°C (0.01°F)
Cu10.01 C 0.1°C (0.1°F)
J 0.4°C
K 0.5°C
T 0.4°C
E 0.4°C
R 1.4°C
S 1.4°C
B 1.4°C
L 0.45°C
U 0.45°C
N 0.6°C
XK 0.4°C
BP 1.1°C
TC Resolution
J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
Specifications
Ramp functions
Source functions
Voltage, current, resistance, frequency, temperature
Ramps
Slow ramp, Fast ramp, 25% step-ramp
Loop power function
Voltage 24 V
Accuracy 0.1
Maximum current
22 mA, short circuit protected
Step functions
Source functions
Voltage, current, resistance, frequency, temperature
Steps
25% of range, 100% of range
Environmental Specifications
Operating temperature -10°C to +55°C
Storage temperature -20°C to +71°C
Operating altitude 3000 m
Safety Specifications
Agency approvals EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Mechanical and General Specifications
Size 200 x 96 x 47 mm (7.9 x 3.8 x 1.9 in)
Weight 650 g (23 oz)
Batteries 4 AA alkaline batteries
Battery replacement Separate battery compartment, accessible without breaking calibration seal
Side port connections Pressure module connector, also used for remote real-time programming

Models

Fluke 726 Precision Multifunction Process Calibrator

Includes:

  • One set of AC72 alligator clips
  • One set of stackable test leads
  • One set of TL75 testleads
  • User manuals
  • NIST- traceable calibration report and data

Resources