Skip to content

Fluke 725 Multifunction Process Calibrator

Key features

  • Calibration of both temperature and pressure transmitters
  • Measure/source/simulate mA loop parameters
  • Measure and source mA, Volts DC, frequency, and resistance
  • Built-in 24 V loop power supply
  • Store and recall frequently used calibration set ups
Category:

Fluke 725 Multifunction Process Calibrator

With the Fluke 725 Multifunction Process Calibrator you can easily source and measure almost any process parameter and test almost any process device. Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.

From the moment you pick it up, the Fluke 725 is ready to perform as a powerful, yet easy-to-use field calibrator. Its simple controls, without menus, make operation easy and the backlit display provides better visibility in poor light. It features a small, streamline shape and a rugged, reliable design that make it ideal for standing up to harsh field conditions.

Versatile troubleshooting tool

Power transmitters during test using the loop supply with simultaneous mA measurement so you can test a transmitter out of circuit and better isolate problems. The 725 can also test a variety of fast pulsed RTD transmitters and PLCs (Programmable logic controllers), with pulses as short as 10 ms.

Source mA with simultaneous pressure measurement to conduct valve and I/P (Input current/Output pressure) tests in a single device. You can also test flow meters with frequency and CPM (counts per minute) functions and perform fast linearity tests with auto-step and auto-ramp features so you can quickly verify the mA signal at different output levels.

Specifications

Measurement Accuracy
Voltage DC
30.000 V
0.02% + 2 counts (upper display)
30.000 V
0.02% + 2 counts (lower display)
100.00 mV
0.02% + 2 counts
-10.00 mV to 75.00 mV
0.025% + 1 count (via TC connector)
Current DC 24.000 mA
0.02% + 2 counts
Resistance
0.0 to 400.0 Ω
0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω
0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω
1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
Frequency
2.0 to 1000.0 CPM 0.05% + 1 count
1.0 to 1100.0 Hz 0.05% + 1 count
1.00 to 10.00 kHz 0.05% + 1 count
Sensitivity
1 V peak-to-peak-minimum
Pressure Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC
100.00 mV
0.02% + 2 counts
10.000 V
0.02% + 2 counts
-10.00 mV to 75.00 mV
0.025% + 1 count (via TC connector)
Current DC
24.000 mA (source)
0.02% + 2 counts
24.000 mA (simulate)
0.02% + 2 counts
Resistance
15.0 to 400.0 Ω
0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω
0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω
1 Ω (excitation current 0.05 to 0.4 mA)
Frequency
2.0 to 1000.0 CPM 0.0005
1.0 to 1100.0 Hz 0.0005
1.00 to 10.00 kHz 0.0025
Waveform
5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure accuracy
NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.8°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution
J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
XK 0.6°C
BP 1.2°C
Source accuracy
NI-120 0.2°C
PT-100 (385) 0.33°C
PT-100 (393) 0.3°C
PT-100 (JIS) 0.3°C
PT-200 (385) 0.2°C
PT-500 (385) 0.3°C
PT-1000 (385) 0.2°C
Resolution 0.1°C
Accuracy stated for 4-wire measurement.
J 0.7°C
K 0.8°C
T 0.8°C
E 0.7°C
R 1.4°C
S 1.5°C
B 1.4°C
L 0.7°C
U 0.75°C
N 0.9°C
Resolution
J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C
XK 0.6°C
BP 1.2°C
Specifications
Ramp functions
Source functions
Voltage, current, resistance, frequency, temperature
Ramps
Slow ramp, Fast ramp, 25% step-ramp
Loop power function
Voltage 24 V
Accuracy 0.1
Maximum current
22 mA, short circuit protected
Step functions
Source functions
Voltage, current, resistance, frequency, temperature
Steps
25% of range, 100% of range
Environmental Specifications
Operating temperature -10°C to +55°C
Storage temperature -20°C to 71°C
Operating altitude 3000 m
Safety Specifications
Agency approvals EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Mechanical and General Specifications
Size 130 x 236 x 61 mm
Weight 0.65 kg
Batteries 4 AA alkaline batteries
Battery replacement Separate battery compartment, accessable without breaking calibration seal
Side port connections Pressure module connector, also used for remote real-time programming

Models

Fluke 725 Multifunction Process Calibrator

Includes:

  • One set of AC72 alligator clips
  • One set of stackable test leads
  • One set of TL 75 testleads
  • User manuals
  • NIST- traceable calibration report and data

Resources

Accessories