Fluke 393 FC CAT III 1500 V True-rms Clamp Meter with iFlex

Key features

  • Measure safely with CAT III 1500 V rated clamp meter
  • Thin jaw for access to cables in crowded combiner boxes
  • Sturdy IP54 rated for outdoor use
  • Work efficiently with dc power measurement, audio polarity and visual continuity
Categories: ,

The 393 FC CAT III 1500 V True-rms Clamp Meter with iFlex is designed for technicians who work in high voltage dc environment: solar arrays, wind power, electric railways, data centers battery banks for uninterruptible power supplies. The clamp will measure up to 1500 V dc, 1000 V ac, and up to 999.9 A dc or ac through the clamp jaw.  The included iFlex flexible current probe extended ac current measurements up to 2500 amps.This clamp has a thin jaw, giving you access to cables in crowded combiner boxes. Test leads are designed with your work in mind, and are also rated to CAT III 1500 V dc.

Other key functions:

  • IP54 rated, ideal for work outdoors on solar arrays and wind power systems
  • DC power measurement, showing readings in kVA
  • Audio Polarity indicator helps prevent accidental miswires
  • Visual Continuity turns provide a bright green light in the display, ideal when working in dark and noisy environments
  • Logging and reporting of test results via Fluke Connect software
  • When measuring ac current the included iFlex flexible current probe gives you unparalleled access to cable is tight spaces.  The iFlex probe can be twisted through extremely small spaces and provide accurate current measurements.

The world’s first CAT III 1500 V true-RMS clamp meter

 

What’s in the box:

  • Fluke 393 FC CAT III 1500 V TRMS clamp meter
  • Test leads, CAT III 1500 V rated, right angle plugs, with safety caps
  • iFlex 18 inch flexible current probe
  • TPAK magnetic hanging strap
  • Premium carrying case
  • 3-year warranty

Technical Specifications

Specifications:
General
Maximum voltage between any Terminal and Earth Ground
AC 1000 V
DC 1500 V
Batteries 2 AA IEC LR6 alkaline
Display Dual display with backlight
Automatic Power Off 20 minutes
Electrical
Accuracy
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits]).
Temperature Coefficients Add 0.1 x specified accuracy for each °C > 28 °C or < 18 °C
AC Current: Jaw
Range 999.9 A
Resolution 0.1 A
Accuracy 2 % + 5 digits (10 Hz to 100 Hz)
2.5 % + 5 digits (100 Hz to 500 Hz)
Crest Factor (50/60 Hz) 2.5 @600.0 A
3.0 @500.0 A
1.42 @999.9 A
Add 2 % for C.F. >2
AC Current: Flexible Current Probe
Range 999.9 A
2500 A
Resolution 0.1 A (≤999.9 A)
1 A (≤2500 A)
Accuracy 3 % RD + 5 digits (10 Hz to 500 Hz)
Crest Factor (50/60Hz) 2.5 @1400 A
3.0 @1100 A
1.42@2500 A
Add 2 % for C.F. >2
Position Sensitivity
Distance from Optimum i2500-10 Flex i2500-18 Flex Error
A 0.5 in (12.7 mm) 1.4 in (35.6 mm) ±0.5 %
B 0.8 in (20.3 mm) 2.0 in (50.8 mm) ±1.0 %
C 1.4 in (35.6 mm) 2.5 in (63.5 mm) ±2.0 %
Measurement uncertainty assumes centralized primary conductor at optimum position, no external electrical or magnetic field, and within operating temperature range.
DC Current
Range 999.9 A
Resolution 0.1 A
Accuracy 2 % RD + 5 digits[1]
[1] When using the ZERO (B) function to compensate for offsets.
AC Voltage
Range 600.0 V
1000 V
Resolution 0.1 V (≤600.0 V)
1 V (≤1000 V)
Accuracy 1 % RD + 5 digits (20 Hz to 500 Hz)
DC Voltage
Range 600.0 V
1500 V
Resolution 0.1 V (≤600.0 V)
1 V (≤1500 V)
Accuracy 1 % RD + 5 digits
mV dc
Range 500.0 mV
Resolution 0.1 mV
Accuracy 1 % RD + 5 digits
Amps Frequency: Jaw
Range 5.0 Hz to 500.0 Hz
Resolution 0.1 Hz
Accuracy 0.5 % RD + 5 digits
Trigger Level 5 Hz to 10 Hz, ≥10 A
10 Hz to 100 Hz, ≥5 A
100 Hz to 500 Hz, ≥10 A
Amps Frequency: Flexible Current Probe
Range 5.0 Hz to 500.0 Hz
Resolution 0.1 Hz
Accuracy 0.5 % RD + 5 digits
Trigger Level 5 Hz to 20 Hz, ≥25 A
20 Hz to 100 Hz, ≥20 A
100 Hz to 500 Hz, ≥25 A
Voltage Frequency
Range 5.0 Hz to 500.0 Hz
Resolution 0.1 Hz
Accuracy 0.5 % RD + 5 digits
Trigger Level 5 Hz to 20 Hz, ≥5 V
20 Hz to 100 Hz, ≥5 V
100 Hz to 500 Hz, ≥10 V
DC Power
Range 600.0 kVA (600.0 V dc range)
1500 kVA (1500 V dc range)
Resolution 0.1 kVA
1 kVA
Accuracy 2 % RD + 2.0 kVA
2 % RD + 20 kVA
Resistance
Range 600.0 Ω
6000 Ω
60.00 kΩ
Resolution 0.1 Ω (≤600.0 Ω)
1 Ω (≤6000 Ω)
0.01 kΩ (≤60.00 kΩ)
Accuracy 1 % RD + 5 digits
Capacitance
Range 100.0 μF
1000 μF
Resolution 0.1 μF (≤100.0 μF)
1 μF (≤1000 μF)
Accuracy 1 % RD + 5 digits
Inrush Trigger Level 5 A
Mechanical
Size (L x W x H) 281 mm x 84 mm x 49 mm
Weight (with batteries) 520 g
Jaw Opening 34 mm
Flexible Current Probe Diameter 7.5 mm
Flexible Current Probe Cable Length
(head to electronics connector) 1.8 m
Environmental
Operating Temperature -10 °C to 50 °C
Storage Temperature -40 °C to 60 °C
Operating Humidity Non condensing (<10°C)
≤90 % RH (at 10 °C to 30 °C)
≤75 % RH (at 30 °C to 40 °C)
≤45 % RH (at 40 °C to 50 °C)
Operating Altitude 2000 m
Storage Altitude 12 000 m
Ingress Protection (IP) Rating IEC 60529: IP54 non-operating
Electromagnetic Compatibility (EMC)
International IEC 61326-1: Portable, Electromagnetic Environment, IEC 61326-2-2 CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.
Korea (KCC) Class A equipment (Industrial Broadcast & Communications Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes.
USA (FCC) 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Safety
General IEC 61010-1, Pollution Degree 2
Measurement IEC 61010-2-032: CAT III 1500 V / CAT IV 600 V
IEC 61010-2-033: CAT III 1500 V / CAT IV 600 V
Wireless Radio
Radio frequency certification FCC ID: T68-FBLE, IC: 6627A-FBLE
Wireless Radio Frequency Range 2400 MHz to 2483.5 MHz
Output Power <100 mW
SIMPLIFIED EU DECLARATION OF CONFORMITY
Hereby, Fluke declares that the radio equipment contained in this Product is in compliance with Directive 2014/53/EU.
The full text of the EU declaration is available at the following Internet address: