Skip to content

Fluke 3130-G2M Portable Pressure Calibrator

Key features

The 3130-G2M Portable Pneumatic Pressure Calibrator is ideal for calibrating pressure transmitters, transducers, gauges and similar devices. The 3130-G2M contains everything you need to generate, control and measure pressure, as well as read the output of the device under test (DUT).

Features

  • Measure and generate pressures from vacuum to 2 MPa (300 psi, 20 bar)
  • Internal pump can generate vacuum to -80 kPa (-12 psi, -0.8 bar) or pressure to 2 MPa (300 psi, 20 bar)
  • Supply pressure connection allowing the use of external gas supply up to 2 MPa (300 psi, 20 bar)
  • Includes variable volume for fine adjustment of pressures
  • Pressure measurement accuracy of 0.025% reading ±0.01% FS
  • Electrical measurement and 24 volt supply for close looped calibrations
  • Measure or generate 4 to 20 mA
  • Measure 0 to 30 V DC
  • Powered by internal, high capacity, NiMH battery or universal ac mains adapter
  • Compatible with Fluke 700P Pressure Modules
Category:

Product overview: Fluke 3130-G2M Portable Pressure Calibrator

Pressure generation and control

The internal, motorized pump provides an easy, efficient alternative to time consuming, tiring hand pumps. For applications that involve filling a large volume with pressure, the 3130-G2M allows for connection to an external gas supply such as compressed plant air. The pressure can be fine-tuned using the variable volume.

Pressure measurement

The 3130-G2M offers an onboard pressure sensor with a full scale of 2 MPa (300 psi, 20 bar) and an accuracy of ±0.025% reading + 0.01% FS (includes precision, one-year stability, uncertainty of the calibration standard, and temperature effects). In addition, the 3130-G2M can be used with the Fluke 700P pressure module series to improve measurement performance across the entire range.

Electrical measurement

The 3130-G2M offers electrical measurement capabilities for calibrating pressure transducers and transmitters, including measurements of 4 to 20 mA or 0 to 30 V dc. In addition, the 3130-G2M provides 24 V dc supply to power the device under test and can generate current in the 4 to 20 mA range.

Portability

The 3130-G2M is built into a ruggedized case with internal storage for power supply, test leads, and fittings. In addition, it includes a rechargeable, NiMH battery, allowing for approximately 50 hours of operation. When using the internal pump, the battery is sufficient to provide up to 100 calibration cycles to 300 psi. The battery is recharged by simply plugging the unit in using the included universal mains adapter.

Automation

In the field, the 3130-G2M can be connected to a personal computer with the Fluke Calibration 700SC Serial Interface Cable Assembly to log data and send instructions using Fluke Calibration COMPASS for Pressure Software. To learn more about the cable assembly and software, contact your local Fluke Calibration sales representative.

Specification

Specifications: Fluke 3130-G2M Portable Pressure Calibrator

Environmental
Operating temperature -10°C to +50°C
Storage Temperature -20°C to +60°C
Power requirements 12 V DC (Universal ac adapter/charger supplied)
Battery Internal 3800 mAh advanced NiMH pack
Operating Time
On full charge Approx 50 hours (Measure only or external air; no pump)
Using internal pump Approx 100 calibration cycles to 300 psi
Physical
Dimensions (L x W x D) 15.25 x 12 x 7 in
Weight ~7 kg (15 lb)
EMI/RFI conformance EN61326:2006 Annex A
Connectors/ports 1/8 in NPT (External supply port and test port)
Included accessories Manual, NIST-traceable certificate, test leads, universal ac adapter/charger
Ranges
Pressure (internal pump) -80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar)
Pressure (external air) 0 to 2 MPa (0 to 300 psi, 0 to 20 bar)
mA 0 to 24.000 mA
Volts 0 to 30.000 V DC
Engineering units psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4°C, mmH2O @ 20°C, cmH2O @ 4°C, cmH2O @ 20°C, inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F, mmHg @ 0°C, inHg @ 0°C
Instrumental Measurement Uncertainty
Pressure 0.025% of reading ±0.01% FS
mA 0.015% of reading ±0.002 mA
Volts 0.015% of reading ±0.002 V
Temperature Effect (All Functions)
No effect on accuracy on all functions from 15°C to 35°C
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C

 

Models

Models: Fluke 3130-G2M Portable Pressure Calibrator

Fluke 3130-G2M

Fluke 3130-G2M Portable Pressure Calibrator
US line cord (standard calibration)

Resources

Resources: Fluke 3130-G2M Portable Pressure Calibrator