General Specifications |
Measuring functions |
3-pole earth ground resistance, 2-pole ac resistance of a conductor, Interference voltage |
Intrinsic error |
Refers to the reference temperature range and is guaranteed for one year |
Measuring rate |
2 measurements/second |
Battery1 |
One 9 volt alkaline (LR61) |
Battery condition |
LO-BAT is displayed if voltage drops below 6.5 V |
Voltages
|
Between jacks H/C2 and E/C1 |
250 Veff maximum (effective voltage) |
Between jacks S/P2 and E/C1 |
250 Veff maximum |
Climatic class |
VDE/VDI 3540 RZ (conforming to KWG as per DIN 40040, 4/87) |
Temperature performance2
|
Working |
-10°C to +50°C (+14°F to +122°F) |
Operating |
0°C to +35°C (+32°F to +95°F) |
Storage |
-20°C to +60°C (+68°F to +140°F) |
Reference |
#ERROR! |
Temperature coefficient |
±0.1% of range per degree Kelvin |
Safety |
IEC/EN 61010-1, 600 V CAT II, pollution degree 2 |
Dimensions |
113 x 54 x 216 mm (4.5 x 2.1 x 8.5 in), including holster |
Weight |
850 g (1.9 lb), including standard accessories, volume approximately 600 cm3 |
Electrical Specifications |
Maximum deviations
|
E1 Influence factor |
Position |
E1 Deviation influence |
0% |
E2 Influence factor |
Supply voltage |
E2 Deviation influence |
0% |
E3 Influence factor |
Temperature E3 |
E3 Deviation influence |
2.30% |
E4 Influence factor |
Serial interference voltage (20 V) |
E4 Deviation influence |
0.60% |
E5 Influence factor |
Probe- and auxiliary probe resistance |
E5 Deviation influence |
10% |
Test voltage |
3.7 kV |
Protection type |
IP 40; IEC/EN 60529 |
Electromagnetic compatibility
|
Emission |
IEC/EN 61326 Class B |
Immunity |
IEC/EN 61326 Annex C |
RE Resistance measurement
|
Measuring method |
Current-voltage measurement with improved cross-talk attenuation, no compensation of measuring lead resistance, with probe (3-pole) or without probe (2-pole), as per IEC/EN 61557–5
|
Open circuit voltage |
23 to 24 V AC |
Short circuit current |
> 50 mA AC |
Measuring frequency |
128 Hz |
Maximum permissible overload |
250 Veff |
Measuring time |
8 seconds (average from when START is pressed) |
Limit input |
Tester retains set value even if instrument is turned off (assuming battery power supply is sufficient) |
Automatic changeover of resolution
|
RH |
< 7 kΩ |
Resolution |
0.01 Ω |
RH |
< 50 kΩ |
Resolution |
0.1 Ω |
RH |
> 50 kΩ |
Resolution |
1 Ω |
Interference voltage display DC + AC
|
Vmax |
30 Veff |
Common mode rejection |
> 80 dB at 50 Hz and 60 Hz |
Ri |
680 kΩ |
Measuring uncertainty |
< 10% for pure AC and DC signals |
Measuring Range |
0.15 Ω to 20 Ω
|
Resolution |
0.01 Ω |
Display range |
0 to 19.99 Ω |
200 Ω
|
Resolution |
0.1 Ω |
Display range |
20 to 199.9 Ω |
2 kΩ
|
Resolution |
1 Ω |
Display range |
200 to 1999 Ω |
Intrinsic Uncertainty |
±(6% of measured value + 5D) |
Operating uncertainty IEC 615573 |
±(18% of measured value + 5D) |